Home page / Product Families / Microbiological Instruments / BacTrac 4300
BacTrac 4300 mikrobiológiai impedancia analizátor
With 64 individual measurement positions and 2 different temperature zones per instrument BacTrac can deal with almost all tasks the microbiologist is familiar with from the classical agar plate.
Measurement principle
Microorganisms are detected in real time via the decrease of the impedance in an AC field. Any impedance change caused by the microbial metabolism is detected.
In the BacTrac devices a proprietary impedance technology is employed. Usind this technology a standard impedance signal ( Media Impedance = M- value) is registered as the reuslt of microbial growth.
During the growth of microbes low- or non charged nutrient molecules are metabolised. This break down of nutrients generates small and highly charged metabolites. The later will result in a change of the impedance of the nutrient media that is registered by the instrument.
The more microbes are present in the sample the sooner results will be available.
HYGENE INDICATORS | PATHOGENES | SPECIAL AREAS |
Total aerobic mesophilic counts (TMAC) | Salmonella | Clostridia |
Enterobacteriaceae | Listeria | Sterility Tests |
Coliforms | Cronobacter | Lactic acid bacteria |
E.coli | Coagulase positive Styphylococci | Osmotolerant Yeasts |
Yeasts and Moulds | Bacillus Cereus | Aerobic sporeformers |
Pseudomonades | Sulfite reducing Clostridia | |
Pseudomonas aeruginosa | Clostridium perfringens | |
Enterococci | Activity testing | |
Desinfectant efficacy testing | ||
ISO 17516 spec. microbes |
Comprises BacMonitor and BacEval measuring and data evaluation software for Windows 7, WIN-8.1 and WIN-10. Traditional Windows menu functionality is easy to understand and operate without extensive training.
Real Time Results - Color Coded Alerts
Database Functions
Specifics